OxySat Pulse Oximeter

The MRI-compatible pulse oximeter OxySat guarantees patient safety by application of fibre-optical sensor systems.

You and OxySat will have certainty on your patients' status.

The OxySat offers you continuous pulse and oxygen saturation values, either numerically or graphically. The pulse oximeter determines saturated hemoglobin within pulsating arterial blood.

Unique:

Plethysmogram for:
• optimum sensor application
• determination of artifacts
• determination of poor circulation

Tesla OxySat and its dual safety system:
• interference-free accurate vital parameters for anesthetists and pediatricians guarantee patient safety.
• interference-free images for radiologists guarantee patient safety.

Dual safety is achieved by high quality standards. Extensive test series and long-term examinations with MR systems of all common field strengthens and sequences. In-house production of the sensor system.

Unique:

Special pressure-reduced light receiver for children and adults. This prevents necrosis during long-term checkups:
• special butterfly sensor system for neonates advantage: suitable for very small extremities.
• transmission by optical fibres advantage: patient safety
• optical fibres are coated with sheathed spiral tube advantage: prevention of footstep damage
• sensors are completely free of metals advantage: patient safety

Oxy Sat Pulse Oximeter
Shown: OxySat Pulse Oximeter (Monitor Only)

When Ordering, Please Specifiy Sensor Type:
03-AN3010 - Neonatal Finger Sensor
03-AN3020 - Adult/Child Finger Sensor

 



 


Other advantages:

- clear display
- individually adjustable alarm parameters
- absolutely metal-free fiber optic sensors
- integrated power pack
- battery operated
- extremely bright plethysmogram
- rugged sensors for everyday clinical life

Item Number: Description:
03-AN3000 OxySat Pulse Oximeter
03-AN3010 Neonatal Finger Sensor (5 meter)
03-AN3020 Child/Adult Finger Sensor (5 meter)
SPECIFICATIONS

 
 
Height 5.1 inches
Depth 10.7 inches
Width 10.2 inches
Weight 14.8 LBS
Operating Temperature Range 41 to 104 degrees, F.
Voltage Range 85 - 265 V at 47- 63 Hz
Battery Operation >240 min.
Positioning - Can be placed as close as 4.5 ft. from center of bore Maximum field strength 20mT
Heart Rate  
 Measuring Range 20BPM - 255 BPM
 Accuracy at 40 - 235 1/min +/- 1.7% of measured value
 Resolution 1 BPM
SpO2  
 Measuring Range 1-100%
 Accuracy at 80 -100% +/- 3% of measured value
 Resolution 1%